FPI AQMS-300 Ozone analyzer measures ambient O3 concentration in ppb level by utilizing UV photometric absorption technology.
The concentration of ambient ozone is proportional to UV light absorption since there is a significant characteristic absorption for ozone on wavelength of 254nm.
Periodical diversion on sample flow passing through the ozone scrubber will generate reference measurement, which is compared with sample measurement to provide stable and representative result.
|
Principe |
UV Photometric |
|
Gamme standard |
Max:0~10ppm Min:0~100ppb (Selectable)
|
|
Zéro bruit |
0.4ppb (RMS) |
|
Bruit de spectre |
<2.5ppb (RMS) |
|
Limite inférieure de détection |
0.8ppb (RMS) |
|
Dérive zéro |
±2ppb/24h |
|
Dérive de la portée |
±5ppb/24h |
|
Linéarité |
<1%F.S. |
|
Précision |
<1% |
|
Temps de réponse |
T90<30s |
| Sortie |
RS232/RS485/Ethernet |
|
Calibrage |
Multi-point Calibrator |
| Puissance requise | 220V AC/50Hz ;
110V AC/60Hz |
Ozone scrubber
The ozone scrubber is filled with MnO2 as catalyst, which will convert O3 to O2 to form reference gas. Meanwhile, the presence of other components remains the same.
Stockage et analyse des données
Les données stockées sont facilement récupérables via le port série ou Ethernet par l'intermédiaire d'un logiciel client PC, ce qui permet aux opérateurs d'effectuer des diagnostics prédictifs et une analyse améliorée des données en suivant les tendances des paramètres.