The concentration of ambient ozone is proportional to UV light absorption since there is a significant characteristic absorption for ozone on wavelength of 254nm.
Periodical diversion on sample flow passing through the ozone scrubber will generate reference measurement, which is compared with sample measurement to provide stable and representative result.

AQMS-300 O3 Analyzer(UV Photometric Absorption Technology) Specifiche tecniche

Principio

UV Photometric

Standard Range

Max:0~10ppm Min:0~100ppb (Selectable)

Rumore zero

0.4ppb (RMS)

Rumore di campata

<2.5ppb (RMS)

Lower Detectable Limit

0.8ppb (RMS)

Deriva zero

±2ppb/24h

Deriva della campata

±5ppb/24h

Linearità

<1%F.S.

Precision

<1%

Tempo di risposta

T90<30s

Uscita

RS232/RS485/Ethernet

Calibration

Multi-point Calibrator

Power Requirement 220V AC/50Hz;

110V AC/60Hz

Applicazioni versatili in tutti i settori

Caratteristiche principali di AQMS-300 O3 Analyzer(UV Photometric Absorption Technology)

  • Accurate direct UV absorption with reference comparison;
  • Compliance with US EPA equivalent method;
  • Various outputs include ethernet and RS232;
  • User friendly interface with large screen;
  • Continuous system diagnosis with alarm;
  • Multi-tasking software allows viewing test variables while operating;
  • Temperature and pressure compensation;

Ozone scrubber
The ozone scrubber is filled with MnO2 as catalyst, which will convert O3 to O2 to form reference gas. Meanwhile, the presence of other components remains the same.

Data storage and analysis
Stored data are easily retrievable through the serial or ethernet port via PC client software, allowing operators to perform predictive diagnostics and enhanced data analysis by tracking parameter trends.

Perché AQMS-300 O3 Analyzer(UV Photometric Absorption Technology) Si distingue

Domande frequenti

Richiedi subito un preventivo

Sbloccare l'analisi di precisione con AQMS-300 O3 Analyzer(UV Photometric Absorption Technology)

Siete pronti a migliorare le vostre operazioni con soluzioni di analisi online leader del settore? Contattateci per ricevere una guida esperta, i prezzi o una demo personalizzata.

Prodotti correlati

Inviate oggi stesso la vostra richiesta

Compilate il modulo sottostante e vi contatteremo al più presto.
Informazioni di contatto
Dettagli del progetto