Casa / Prodotto / Analizzatore di CO AQMS-400

Analizzatore di CO AQMS-400

FPI AQMS-400 carbon monoxide (CO) analyzer measures ambient CO concentration by employing nondispersive infrared with gas filter correlation method technology.

Download e risorse

Richiedi subito un preventivo
Descrizione

Panoramica del prodotto

FPI AQMS-400 carbon monoxide (CO) analyzer measures ambient CO concentration by employing nondispersive infrared with gas filter correlation method technology.

Principio

Infrared energy emitted by light source is passed through gas chamber containing the air sample, and the quantitative absorption of energy by CO in the sample cell is measured by corresponding detector.

Analizzatore di CO AQMS-400 Specifiche tecniche

Standard Range
Max:0~1000ppm Min:0~50ppm (Selectable)
Principio NDIR
Rumore zero ≤0.02ppm (RMS)
Rumore di campata ≤0.25ppm (RMS)
Lower Detectable Limit ≤0.04ppm (RMS)
Display Digital
Deriva zero <0.1ppm/24h
Deriva della campata ±0.5ppm/24h
Linearità <1%F.S.
Precision <1%
Tempo di risposta T90<60s
Sample Flow Rate (800±80)sccm
Data Transmission 2 channel analog (4~20) mA;

12 digital input/output;

4-way relay output;

Communication RS232/RS485/Ethernet
Temperatura di esercizio US EPA Specification 20~30°C

Actual applicable:-5~55°C

Operating Humidity 0~95%RH
Calibration Multi-point calibrator
Power Requirement 220V AC/50Hz;

110V AC/60Hz

Dimensions and Weight 178(H)x432(W)x597(D)mm, 28kg

Applicazioni versatili in tutti i settori

Caratteristiche principali di Analizzatore di CO AQMS-400

  • 14 meters optical path to perform high reliability.
  • Five years guarantee on GFC wheel.
  • Compliance with US EPA reference method.
  • Various outputs include ethernet and RS232.
  • User friendly interface with large screen.
  • Diagnosi continua del sistema con allarme.
  • Multi-tasking software allows viewing test variables while operating;
  • Temperature and pressure compensation.
  • Internal data logging with 1 min to 365 day multiple averages.

GFC
GFC (Gas filter correlation) technology is utilized to remove interference caused by moisture and other backgrounds.

Two gas filled chambers are mounted on a rotating disc, which pass through an IR beam alternately. The measure chamber is filled with nitrogen while the reference chamber is filled with high concentration CO. IR beam then passes through the sample gas cell. The difference in absorbance is measured and provides a direct output of the gas concentration.

Data storage and analysis
Stored data are easily retrievable through the serial or ethernet port via PC client software, allowing operators to perform predictive diagnostics and enhanced data analysis by tracking parameter trends.

Perché Analizzatore di CO AQMS-400 Si distingue

Domande frequenti

Richiedi subito un preventivo

Sbloccare l'analisi di precisione con Analizzatore di CO AQMS-400

Siete pronti a migliorare le vostre operazioni con soluzioni di analisi online leader del settore? Contattateci per ricevere una guida esperta, i prezzi o una demo personalizzata.

Prodotti correlati

Inviate oggi stesso la vostra richiesta

Compilate il modulo sottostante e vi contatteremo al più presto.
Informazioni di contatto
Dettagli del progetto