Metals analyzer creatively combined with world leading argon circulation technology enables the analysis of Fe, Al, Cu and other bases materials. It is an ideal solution to fast and accurate quanti-tative analysis near the furnace, metal material quality control and scientific research.
Metals analyzer creatively combined with world leading argon circulation technology enables the analysis of Fe, Al, Cu and other bases materials. It is an ideal solution to fast and accurate quanti-tative analysis near the furnace, metal material quality control and scientific research.
Principle
M5000 series adopt the most advanced CMOS sensor technology, with a sturdy and stable constanttemperature dual-optical chamber structure, synthesis pulse digital source, and an intelligent logiccorrection program. Ensure that the analysis results are efficient, accurate, and reliable.
| Optical System | Paschen-Runge Mounting |
| Power Supply | Single Phase (220±20)VAC, (50±1)Hz |
| EMC | IEC61000-4-2, IEC61000-4-4, IEC61000-4-5 |
| Maximum Excitation Power | 400VA |
| Average Standby Power | 50VA |
| Sparking Power Supply | Synthesis Pulse Digital Source |
| Spark Pulse | 20~230V |
| Arc Excitation Pulse | 20~60V |
| Working Temperature | (10~30)°C |
| Working Humidity | 20%~80% |
| Argon Purity Requirement | 100.00% |
| Argon Inlet Pressure | 0.5MPa |
| Argon Flow | Standby Flow 0.1L/min; Excitation Flow 3.5L/min |
| Size/Weight | L726 x W622.5 x H546.5mm / 80kg |