AQMS-550 is equipped with an internally mounted catalytic converter set at 315°C to convert H2S to SO2. Bypassing the sample through an SO2 scrubber which removes any SO2, the H2S is converted to SO2 using this internal H2S converter and measured using the pulsed fluorescence technique. AQMS-550 H2S analyzer adopts UV fluorescence principle with an internal thermal catalytic converter to measure H2S in ambient air at levels commonly required.
AQMS-550 is equipped with an internally mounted catalytic converter set at 315°C to convert H2S to SO2. Bypassing the sample through an SO2 scrubber which removes any SO2, the H2S is converted to SO2 using this internal H2S converter and measured using the pulsed fluorescence technique.
Principle
AQMS-550 H2S analyzer adopts UV fluorescence principle with an internal thermal catalytic converter to measure H2S in ambient air at levels commonly required.
|
Principle |
UV Fluorescence |
|
Range |
SO2 : Min: 0~50ppb Max: 0~20ppm H2S : Min: 0~50ppb Max: 0~10ppm |
| Lower Detection Limit |
<0.4ppb or 0.5 % of reading |
|
Zero Noise |
<0.2ppb |
|
Span Noise |
<0.5%F.S. |
|
Zero Drift |
<1ppb/24 hours <5ppb/7days |
|
Span Drift |
<1%F.S. |
|
Response Time |
<150 second |
|
Output |
4~20mA, RS232, RS485, Ethernet |